Climate Change Data Portal
DOI | 10.1007/s11581-015-1454-7 |
Monochloramine-sensitive amperometric microelectrode: optimization of gold, platinum, and carbon fiber sensing materials for removal of dissolved oxygen interference | |
Lee, Woo Hyoung1; Wahman, David. G.2; Pressman, Jonathan. G.2 | |
发表日期 | 2015-09-01 |
ISSN | 0947-7047 |
卷号 | 21期号:9页码:2663-2674 |
英文摘要 | Monochloramine electrochemical determination in an aqueous system using newly fabricated gold and platinum microelectrodes was investigated to optimize sensor operation and to eliminate dissolved oxygen (DO) interference during monochloramine measurements. Carbon fiber microelectrodes were also compared for reference purposes. Gold and platinum microelectrodes exhibited no oxygen interference during monochloramine measurement and provided a linear relationship when operated at +150 and +300 mV vs. Ag/AgCl over a wide concentration range (0-4.2 mg Cl-2/L), respectively. The carbon fiber microelectrode with 7-mu m tip diameter was not sufficiently sensitive to monochloramine concentrations for detailed study. The baseline signal of both gold and platinum microelectrodes (i.e., signal without monochloramine) was near zero. With the same geometric tip diameter (20-mu m tip diameter), gold microelectrodes resulted in better amperometric electrode response to monochloramine than platinum microelectrodes; gold microelectrodes had a higher sensitivity (52 +/- 0.7 vs. 18 +/- 0.07 pA/[mg Cl-2/L]) and lower detection limit (0.12 +/- 0.013 vs. 0.33 +/- 0.10 mg Cl-2/L), resulting in gold as the preferred microelectrode material. The developed gold microelectrode will allow accurate in situ monochloramine determination in biofilm while eliminating the confounding effects of oxygen interference. |
英文关键词 | Microelectrode;Monochloramine;Oxygen interference;Gold;Platinum;Carbon fiber |
语种 | 英语 |
WOS记录号 | WOS:000360088500033 |
来源期刊 | IONICS
![]() |
来源机构 | 美国环保署 |
文献类型 | 期刊论文 |
条目标识符 | http://gcip.llas.ac.cn/handle/2XKMVOVA/58459 |
作者单位 | 1.Univ Cent Florida, Dept Civil Environm & Construct Engn, Orlando, FL 32816 USA; 2.US EPA, Natl Risk Management Res Lab, Cincinnati, OH 45268 USA |
推荐引用方式 GB/T 7714 | Lee, Woo Hyoung,Wahman, David. G.,Pressman, Jonathan. G.. Monochloramine-sensitive amperometric microelectrode: optimization of gold, platinum, and carbon fiber sensing materials for removal of dissolved oxygen interference[J]. 美国环保署,2015,21(9):2663-2674. |
APA | Lee, Woo Hyoung,Wahman, David. G.,&Pressman, Jonathan. G..(2015).Monochloramine-sensitive amperometric microelectrode: optimization of gold, platinum, and carbon fiber sensing materials for removal of dissolved oxygen interference.IONICS,21(9),2663-2674. |
MLA | Lee, Woo Hyoung,et al."Monochloramine-sensitive amperometric microelectrode: optimization of gold, platinum, and carbon fiber sensing materials for removal of dissolved oxygen interference".IONICS 21.9(2015):2663-2674. |
条目包含的文件 | 条目无相关文件。 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。